129
Mahr GmbH • Digimar 817 CLM
10.1
Software update
10
Additional functions
Description / Sequence
Symbols / Pictures
Procedure
–
Connect zero modem cable to the COM 1 interface on the PC and
the RS232 OUT interface on the height measuring instrument
Note:
When the COM 1 interface is occupied, select another PC
–
Save both the Eprom-Update-Software and the text file
(e.g. DX1_Rev. 1.00-02.txt) on the PC
(Revision status can be updated)
–
Open Eprom-File with a double left mouse click
–
The Eprom-Update-Software opens
Note:
Make sure that the correct COM interface has been specified and
„Others“ is highlighted
–
Chose „Select file“.
Requirements for a Software update:
Zero modem cable
Order no. 7024634
Eprom-Update-Software (RS232 Flash programmer for Main program of
Mahr) Version V 2.00.3
- made by Mahr.
highlight
–
Click on the actual revision and open. The program loads the text
file and is ready for transmission