38
Mahr GmbH • Digimar 817 CLM
4.1.3
Calibrate a probe with a ledge
–
Press the probe calibration key
–
Press the
„Calibrate with a ledge“
key
The probe automatically travels to the center of the groove of the
setting standard
–
Position the probe below the ledge, the probe moves upwards
and
makes contact
–
Position the probe above the ledge, the probe moves downwards
and makes contact
–
Position the probe for a second time below the ledge, the probe
moves upwards and makes contact
–
Position the probe for a second time above the ledge, the probe
moves downwards and makes contact
Note:
When calibrating with a ledge, it is advisable to use a disc probe.
Attention:
The probe must be manually positioned according to the diameter
above the ledge!
4.1.2.1
Twin probe, change the axis
–
Press the probe calibration key
–
Select horizontal or vertical probe
Description / Sequence
Symbols / Pictures
Note:
The active probe will be shown in the top right hand corner of the
display with either an V=Vertical or H=Horiziontal