Mahr GmbH • Digimar 817 CLM
4.1.2 Calibrate a twin probe
–
Press the
„calibration“
key
–
Press the function key
„Twin probe“
.
The probe automatically travels the center of the groove of the
setting standard
–
Slide the setting standard, so that the probe is inside the groove.
The groove will automatically be measured twice
–
Position the vertical probe above the ledge
–
Press the function key
„Calibrate vertical probe“
.
The measured value is automatically accepted
–
Position the horizontal probe above the ledge
–
Press the function key
„Calibrate horizontal probe“
.
The measured value is automatically accepted
–
The probe constant of the horiziontal probe and the distance
between the horiziontal and vertical probes will be displayed.
The calibration procedure is complete
–
Select horizontal or vertical probe
Description / Sequence
Symbols / Pictures