User Manual LW120
SAMPLE CLOCK
Sources:
Internal
ange
100 mS/s to 100 MS/s
Resolution
7 digits limited by 1
µ
S/s (10 digits using GPIB programming)
Accuracy
Same as reference
Stability
Same as reference
Reference
Standard
≥
0.01% (100 ppm oscillator) initial tolerance over a 19
°
C to
29
°
C temperature range; 100ppm/
°
C below 19
°
C and above
29
°
C; 100ppm/year aging rate
Option
≥
0.0001% (1 ppm TCXO) initial tolerance over a 19
°
C to
29
°
C temperature range; 1ppm/
°
C below 19
°
C and above
29
°
C; 1ppm/year aging rate
External
10 MHz TTL, 50%
±
2% duty cycle
External
From rear panel BNC, DC to 100 MHz
Level
ECL 100K compatible
SAMPLE CLOCK MODULATION
FM – Built-in Standard Waveforms
Description
Sample clock can be frequency modulated by internal
waveforms that are resident in internal memory (fixed
waveforms).
Modulation Source
mp
Modulation Frequency Range
1 mHz to 100 KHz
Resolution
7 digits
Accuracy
0.1%
Modulating Frequency Distortion
<0.1%
Deviation Range
100mS/s to 100 MS/s
Advance
Automatic, triggered, gated or software command
Marker
Output and Level
Same as SYNC output.
Position
Programmable for selected frequency
M – Downloaded Arbitrary Waveforms
Description
Sample clock can be frequency modulated arbitrary
waveforms that are downloaded by the user (user
waveforms.
Modulation Source
User waveform, any shape, 10 to 20000 waveform points
Modulation Sample Clock Range
1 mS/s to 2 MS/s
Resolution
7 digits
Accuracy
0.1%
Modulating Frequency Distortion
<0.1%
Sample Clock Deviation Range
100mS/s to 100 MS/s
Advance
Automatic, triggered, gated or software command
Marker
Output and Level
Same as SYNC output.
Position
Programmable for selected frequency
Waveform Download
Through RS232, or DMA channel through GPIB
R
Internal sine square, triangle and ra
F
Specifications A-2