User Manual LW120
3. Activating filters when the LW120 is set to output the built-in sine
waveform, or activating the built-in sine waveform when one of the
LW120 filters is turned on. Corrective action: If in sine, select
another function and activate the filter(s).
4. Activating burst mode when the LW120 is set to sequence mode, or
activating sequence mode when the LW120 is set to burst mode.
Corrective action: Remove the LW120 from burst or sequence and
then selected the desired mode.
5. Changing operating mode from triggered to continuous when the
LW120 is set to single sequence advance, or changing the
operating mode from continuous to triggered when the LW120 is
set to automatic sequence advance mode. Corrective action:
Observe the LW120 advance mode while setting sequence
advance.
-222,”Data out of range”. Parameter data, which followed a specific
header, could not be used because its value is outside the valid range
defined by the generator.
-224,”Illegal parameter value”. A discrete parameter was received
which was not a valid choice for the command. An invalid parameter
choice may have been used.
-300,”Device-specific-error”. This is the generic device-dependent
error for the instrument when it cannot detect more specific errors. A
device- specific error as defined in IEEE-488.2 has occurred.
-311,”Memory error”. Indicates that an error was detected in the
instrument’s memory.
-350,”Queue Overflow”. The error queue is full because more than 30
errors have occurred. No additional errors are stored until the errors
from the queue are removed. The error queue is cleared when power
has been shut off, or after a *CLS command has been executed.
-410,”Query INTERRUPTED”. A command was received which sends
data to the output buffer, but the output buffer contained data from a
previous command (the previous data is not overwritten). The output
buffer is cleared when power is shut off or after a device clear has
been executed.
Remote Programming Reference 4-68