User Manual LW120
OUTPut
m
This subsystem controls the characteristics of the output. It controls
Subsyste
output, source and position. Optional nodes were omitted from these
commands. Factory defaults after *RST are shown in bold typeface.
Parameter low and high limits are given where applicable.
arameter
Form
Default,
Low
Limit,H
filter type and filter cutoff frequency. It is also used to control the sync
Keyword P
igh
Limit
:OUTPut (?)
{OFF|ON|0|1}
OFF
:FILTer (?)
{NONE|25M|50M|ALL}
NONE
:SYNC
(?) {OFF|ON|0|1}
OFF
:SOURce (?)
{BIT|LCOMplete}
BIT
:POSition (?)
<position>
0
,0,999999
:WIDTh (?)
<width>
4
,4,100000
OUTPut{OFF|ON|0|1}
Purpose
This command will turn the LW120 output on and off.
•
ON or 1 will set the output on
•
OFF or 0 will set the output off
Parameter type
Discrete
OUTPut?
Response and default
The LW120 will return 1 if the output is on, or 0 if the output is off. Default value is 0.
OUTPut:FILTer{NONE|25M|50M|ALL}
Purpose
This command will select which filter is connected to the LW120 output. Observe the following restrictions
when you try to use this command:
1) Filter selection is not available when the instrument is set to output the standard sine waveform. In fact,
the default waveform shape is sine. Therefore, filter selection will be available for use only after you
select a different waveform, or change the output mode to use.
ilters are placed before the output amplifier. Therefore, do not expect the filters to remove in-band
amplifier harmonics and spurious.
2) F
•
the output path
NONE will disconnect all filters from
•
25M will connect 25 MHz low pass elliptic filter
•
50M will connect 50 MHz low pass elliptic filter
•
ALL will connect both the 25 MHz and the 50 MHz filters in series
Parameter type
Discrete
Remote Programming Reference 4-34