User Manual LW120
SYSTem
Subsystem
The system-related commands are not related directly to waveform
generation but are an important part of operating the LW120. These
commands can reset or test the instrument, or query the instrument
for system information.
Keyword Parameter
Form
Default,
Low
Limit,High
Limit
:SYSTem
:ERRor?
:VERSion?
:SAVE
<memory#>
1
,1,10
:RECall
<memory#>
1
,1,10
RESet
*RST
TEST?
*TST?
*IDN?
SYSTem:ERRor?
Purpose
This query will interrogate the LW120 for programming errors.
Response to query
The LW120 will return error code. Error messages are listed later in this manual.
SYSTem:VERSion?
Purpose
This query will interrogate the LW120 for SCPI version.
Response to query
The LW120 will return 1999.0.
SYSTem:SAVe
<memory#)
Purpose
This command will save the entire front panel setting. This command will not save arbitrary memory and
segment partitions. It will also not save sequence table data.
•
<memory#> will select the memory cell where front panel settings will be stored.
Parameter type and range
Integers from 1 through 10
SYSTem:RECall
<memory#)
Purpose
This command will recall and update front panel setting from parameters stored with the SYST:SAVe
command. This command will not restore arbitrary memory and segment partitions. It will also not restore
sequence table data.
Remote Programming Reference 4-58