User Manual LW120
<MAXimum>
10 in units of volts
VOLTage?
Response to query version
The LW120 will return the present amplitude value. The returned value will be in standard scientific format
(for example: 100mV would be returned as 100E-3 – positive numbers are unsigned). Default value is 5.
VOLTage:OFFSet<offs>
Purpose
This command programs the amplitude offset of the output waveform. The offset is calibrated when the
source impedance is 50
Ω
.
•
<offs> sets the amplitude offset
Parameter type
Numeric
Parameter range
<offs>
-4.5 to +4.5 in units of volts
VOLTage:OFFSet?
Response and default
The LW120 will return the present offset value. The returned value will be in standard scientific format (for
example: 100mV would be returned as 100E-3 – positive numbers are unsigned). Default value is 0.
SINusoid:PHASe<phase>
Purpose
This command programs start phase of the standard sine waveform. This command has no affect on
arbitrary waveforms.
•
<phase> sets the start phase
Parameter type
Numeric
Parameter range
<phase>
0 to 360 in units of degrees
SINusoid:PHASe?
Response and default
The LW120 will return the present start phase value. Default is 0.
TRIangle:PHASe<phase>
Purpose
This command programs start phase of the standard triangular waveform. This command has no affect on
arbitrary waveforms.
•
<phase> sets the start phase
Parameter type
Numeric
Remote Programming Reference 4-29