Appen
SPECIFICATION
Configuration
Output Channels
2, semi-independent
Inter-Channel Control
Leading Edge Offset
Description
Channel 2 edge trails channel 1 edge by a programmable
number of points.
Offset Units
Waveform points
Range
0 to 999999 points
Resolution and Accuracy
1 point, or 1 sample clock period of channel 2
Initial
Skew
<
±
2 ns, with sclk divider = 1; <
±
3 ns, with sclk divider > 1
Channel 2 Sample Clock Divider
Description
The sample clock source is common to both channels 1 and
2. The sample clock for the slave channel can be divided
down coherently from 1 to 64k.
Range
1 to 65,535
Resolution
1
Inter-Channel
Dependency
Separate controls
Output on/off, amplitude, AM, offset, standard waveforms,
user waveforms, user waveform size, sequence table,
dix A
channel 2 clock divider, trigger start phase, breakpoints
Common Cont
source, trigger modes,
tput, FM, FSK, sweep,
arm start/stop
Multiple Instrument synchronization
Description
Multiple instruments can be connected together and
synchronized to provide multi-channel synchronization.
Phase (leading edge) offset between master and slave units
is programmable.
Leading Edge Offset
Description
Leading edge of master output trails the leading edge of the
slave output by a programmable number of points. Each
slave can be programmed to have individual offset
Phase Units
Waveform points
Range
0 to 999999 points
Resolution and Accuracy
4 point
Initial
Skew
<
±
15 ns, depending on cable length and quality, typically with
1 meter coax cables
rols
Sample clock, frequency, reference
sequence advance mode, SYNC ou