User Manual LW120
Basic LW120
Option LNPC Installed
Range
Index
Amplitude Range
Offset Range
Offset
Resolution
Offset Range
Offset
Resolution
Window
1
10.00V to 3.162V
±
4.500V
1mV
±
4.50000V 10
µ
V
±
5V
2
3.161V to 1.000V
±
4.500V
1mV
±
4.50000V 10
µ
V
±
5V
3
999.9mV to 316.2mV
±
450.0mV
0.1mV
±
450.000mV 1
µ
V
±
500mV
4
316.1mV to 100.0mV
±
450.0mV
0.1mV
±
450.000mV 1
µ
V
±
500mV
5
99.99mV to 31.62mV
±
45.00mV
0.01mV
±
45.0000mV 0.1
µ
V
±
50mV
6
31.61mV to 10.00mV
±
45.00mV
0.01mV
±
45.0000mV 0.1
µ
V
±
50mV
Table 3-4. Amplitude – Offset Inter-Relation
Notes
1. While setting amplitude and offset, note that the
Amplitude Range defines the Range Index. Legal
offset values may be set using the appropriate Range
Index. Front panel setting will automatically detect
the appropriate range and will not allow error
conditions. If you program amplitude and offset from
remote and use two different Range Indexes, an error
will occur.
2. Offset fine is an exception. You may exceed the
window range with the offset fine and no error will
result.
Observe Figure 3-7 and modify amplitude and offset using the
following procedure.
1. Press the AMPLITUDE button
2. Select Ampl CH1, Offs CH1, or Offs Fine CH1 parameter. Press
More to access channel 2 amplitude and offset parameters.
3. Move the cursor from left to right until it appears under the digit
you want to change
4. Use the dial to modify the parameter. The output waveform is
updated immediately as you rotate the dial.
Using The Instrument 3-9