Check humidity monitoring
Deadly risk!
Working on open, live equipment!
• Do not operate the equipment with damp hands!
• Caution: live components!
Check switching function
Deadly risk!
Switch off operating voltage!
Risk of injury!
Prior to operation, depressurise the equipment by opening the safety valve "high pressure" (25)!
• Close shut-off valve (
9
).
• Unscrew blind cover (
83
) from the pressure limiting valve
(
34
).
•
Undo connecting hose (
85
) between compressor hose and
solenoid valve Y2 at solenoid valve Y2 (
32
).
• Attach the testing hose supplied (
86
) to connecting hose
(
85
) and test nipple (
83
) of the pressure limiting valve (
34
),
in order to bypass the air dryer.
• Open safety valve "high pressure" (
25
) and and discharge
tank. Close safety valve "high pressure" (
25
) after this.
•
Switch on operating voltage:
• after a few minutes at least one of the red LEDs of the 8
digit humidity display (
39
) (7. and 8. digit) illuminates on the
Minidisplay (
14
) as well as the red signal-LED „
F
“ (
48
) on
the MFR (
73
) and the system will be shut down and
85 32
83
34
86 25
9
• the red Signal output-LED „
A
“ (
64
) illuminates on the MFR.
•
If the humidity shut-down process is not executed, discharge tank again by opening safety valve "high
pressure" (
25
), in order to start a new compressor run.
•
Close safety valve "high pressure" (
25
) again.
Restore operating status
Deadly risk!
Switch off operating voltage!
Risk of injury!
Prior to operation, depressurise the equipment by
opening the safety valve "high pressure" (25)!
• Undo and remove test hose (
86
) from pressure limiting valve (
34
) and connecting hose (
85
).
• Screw blind cover (
83
) on the pressure limiting valve (
34
) again.
• Screw connecting hose (
85
) to solenoid valve Y2 (
32
).
•
Switch on operating voltage.
If the compressor fails to start, allow the equipment to "run dry":
Deadly risk!
Electro specialists only are allowed to operate the buttons of the MFR.
Other users must use the Minidisplay to operate the system.
• Unscrew connecting block cover (
11
).
• Open safety valve "high pressure" (
25
).
• Press button „
F-Off
“ (
44
) on the Minidisplay or (
67
) on the
MFR.
•
The compressor starts.
• The red LED „
A
“ (
41
) on the Minidisplay turned off.
•
The green Signal-LED „
F-on
“ (
43
) on the Minidisplay and (
54
)
in the MFR are turned off = no humidity monitoring.
•
at least one of the red LEDs of the 8 digit humidity display (
39
)
(7. and 8. digit) illuminate.
RTS 1000D
25
Содержание RTS 1000D
Страница 43: ...Page 43 Compressor VD 7 2 RTS 1000D and VD 28 2 RTS 2600D B...
Страница 45: ...Page 45 Compressor VD 50 4 RTS 5200D...
Страница 57: ...Compressor VD 7 2 and VD 28 2 Page 57...
Страница 59: ...Page 59 Compressor VD 50 4...
Страница 61: ...Page 61 Pneumatics diagram RTS 1000D RTS 2600D and RTS 5200D no 073445 072...
Страница 63: ...Page 63 Circuit and wiring diagram RTS 1000D RTS 2600D und RTS 5200D no 073659 000...
Страница 65: ...Page 65 Circuit and wiring diagram RTS 1000D RTS 2600D und RTS 5200D no 073586 000...
Страница 66: ...Page 66...
Страница 67: ...Page 67...