![Lake Shore 648 Скачать руководство пользователя страница 88](http://html1.mh-extra.com/html/lake-shore/648/648_user-manual_670640088.webp)
78
CHAPTER 6:
Service
Model 648 Electromagnet Power Supply
6.11.4 Power Supply
Water Connectors
This section describes the flow switch and the water control valve.
6.11.4.1 Flow Switch
The flow switch connector provides the means to connect the mandatory water flow
switch (closed during flow) to protect the Model 648 from loss of water flow. When
water is flowing, pins 1 and 2 will be closed, allowing normal operation.
Flow switch connector
Power supply water flow
switch connector details
Pin
Name
1
Flow switch (com)
2
Flow switch
6.11.4.2 Water Control Valve
The water control valve connector provides the means to connect a water control
valve. Pins 1 and 2 supply 24 VAC at 1 A for operation of a water control valve.
Water control valve connector
Power supply water control
valve connector details
Pin
Name
1
Power supply (water valve A)
2
Power supply (water valve B)
FIGURE 6-9
TABLE 6-7
FIGURE 6-10
TABLE 6-8