6.3.3 Intermittent Lockups
71
|
www.lakeshore.com
6.5 Line Voltage
Selection
The Model 648 may be configured for seven basic AC power configurations: 200 VAC,
208 VAC, 220 VAC, 230 VAC, 380 VAC, 400 VAC, and 415 VAC. Proper voltage selection
must be made before connection to the power mains. Each configuration requires the
appropriate wiring within the power wiring access panel on the rear of the
instrument. Nominal line voltages and appropriate selections are shown in
TABLE 6-1. Refer to section 3.4.1 for further details. See FIGURE 6-2 for general
locations.
Model 648 rear panel : Left: Bottom half (shown with wiring cover removed); Right: Top half
Use the following procedure to change the instrument line voltage.
To avoid potentially lethal shocks, turn off the power supply and disconnect it from AC
power before performing this procedure.
1. Identify the power wiring access panel on the rear of the Model 648.
2. Remove the perimeter screws holding the power wiring access panel.
3. Observe the four voltage-selection wires held by the clear plastic wiring guide.
(FIGURE 6-3).
4. Loosen the screw terminals presently holding the four wires.
5. Relocate the four wires (using the wiring guide) to the desired voltage position.
Use TABLE 6-1 to determine the correct voltage position.
6. Place the wires into the appropriate screw terminals and tighten.
7. Verify solid, tight screw connections to these four wires.
8. Replace the power wiring access panel using all perimeter screws.
9. Verify the voltage indicator in the window of the power wiring access panel.
FIGURE 6-2