5.4 Command Summary
67
|
www.lakeshore.com
RDGI?
Current Output Reading Query
Input
RDGI?[term]
Returned
<current>[term]
Format
+nn.nnnn
<current>
Actual measured output current
RDGV?
Output Voltage Reading Query
Input
RDGV?[term]
Returned
<voltage>[term]
Format
+n.nnnn
<voltage>
Actual output voltage measured at the power
supply terminals.
RSEG
Ramp Segments Enable Command
Input
RSEG <enable>[term]
Format
n[term]
<enable>
Specifies if ramp segments are to be used: 0=Disabled,
1=Enabled.
Remarks
Ramp segments are used to change the output current ramp rate based on the output
current. Ramp segments need to be setup first using the RSEGS command.
RSEG?
Ramp Segments Enable Query
Input
RSEG?[term]
Returned
<enable>[term]
Format
n (refer to command for description)
RSEGS
Ramp Segments Parameters Command
Input
RSEGS<segment>,<current>,<rate>[term]
Format
n, +nnn.nnn, +n.nnnn [term]
<segment>
Specifies the ramp segment to be modified: 1 to 5
<current>
Specifies the upper coutput current setting that will use
this segment: 0.0000-+135.100 A.
<rate>
Specifies the rate at which the current will ramp at when
the output current is in this segment: 0.0001 to 50.000 A/s.
Remarks
Ramp segments are used to change the output current ramp rate based on the output
current. The ramp segment feature needs to be turned on using the RSEG command.
RSEGS?
Ramp Segments Parameters Query
Input
RSEGS?<segment>[term]
Returned
<current>, <rate>[term]
Format
+nnn.nnn, +n.nnnn (refer to command for description)
SETI
Output Current Setting Command
Input
SETI<current>[term]
Format
±nnn.nnn
<current>
Specifies the output current setting: 0.000 - ±135.100 A.
Remarks
Sets the current value that the output will ramp to at the present ramp rate. The set-
ting value is limited by LIMIT.