3.4.5 Power Input Terminals
23
|
www.lakeshore.com
3.4.5 Power Input
Terminals
The Model 648 requires a 4 conductor power cable (non-detachable). The input to the
Model 648 is wired in a delta configuration, but will operate from a delta or wye
source. If operating from a wye source, the neutral line (N) is not used.
The ground (the green/yellow ground terminal) connects the instrument chassis to the
electrical ground (safety ground), and is required to prevent fault conditions which may
be hazardous to operating personnel. In no case should the safety ground line be omitted.
In no case should a neutral line be used as a safety ground.
All wiring must comply with the code requirements of the locality in which the
instrument is installed. FIGURE 3-6 shows typical input wiring. The wire ferrules
shown are not required, but are recommended to prevent stray strands from shorting
to adjacent terminals. For safety, the ground wire must be longer than the phase
conductors, as showin in FIGURE 3-6
Typical input wiring
Ground
Line input
Line input
Ground
FIGURE 3-6