CHAPTER 6:
Service
Model 648 Electromagnet Power Supply
6.3 IEEE Interface
Troubleshooting
This section provides IEEE interface troubleshooting for issues that arise with new
installations, old installations, and intermittent lockups.
6.3.1 New Installation
1. Check the instrument address.
2. Always send a message terminator.
3. Send the entire message string at one time including the terminator.
4. Send only one simple command at a time until communication is established.
5. Be sure to spell commands correctly and use proper syntax.
6. Attempt both Talk and Listen functions. If one works but not the other, the hard-
ware connection is working, so look at syntax, terminator, and command format.
6.3.2 Existing
Installation No Longer
Working
1. Power the instrument off, then on again to see if it is a soft failure.
2. Power the computer off then on again to see if the IEEE card is locked up.
3. Verify that the address has not been changed on the instrument during a
memory reset.
4. Check all cable connections.
6.3.3 Intermittent
Lockups
1. Check cable connections and length.
2. Increase the delay between all commands to 50 ms to make sure the instrument
is not being overloaded
6.4 Fuse Drawer
The fuse drawer supplied with the Model 648 holds the instrument line fuses and line
voltage selection module. The drawer holds two 10.30 mm × 38 mm (0.41 in × 1.5 in)
time delay fuses. It requires two good fuses of the same rating to operate safely.
Refer to section 6.7 for instructions to exchange fuses.
Fuse drawer
FIGURE 6-1