![Lake Shore 648 Скачать руководство пользователя страница 76](http://html1.mh-extra.com/html/lake-shore/648/648_user-manual_670640076.webp)
66
CHAPTER 5:
Computer Interface Operation
Model 648 Electromagnet Power Supply
MODE?
IEEE Interface Mode Query
Input
MODE?[term]
Returned
<mode>[term]
Format
n (refer to command for description)
OPST?
Operational Status Query
Input
OPST? [term]
Returned
<bit weighting> [term]
Format
nnn
Remarks
The integer returned represents the sum of the bit weighting of the operational sta-
tus bits. Refer to section 5.2.6.2 for a list of operational status bits.
OPSTE
Operational Status Enable Command
Input
OPSTE <bit weighting> [term]
Format
nnn
Remarks
Each bit has a bit weighting and represents the enable/disable mask of the corre-
sponding operational status bit in the Operational Status Register. This determines
which status bits can set the corresponding summary bit in the Status Byte Register.
To enable a status bit, send the command OPSTE with the sum of the bit weighting for
each desired bit. Refer to section 5.2.6.2 for a list of operational status bits.
OPSTE?
Operational Status Enable Query
Input
OPSTE?[term]
Returned
<bit weighting> [term]
Format
nnn (Refer to section 5.2.6.2 for a list of operational status bits)
OPSTR?
Operational Status Register Query
Input
OPSTR? [term]
Returned
<bit weighting> [term]
Format
nnn
Remarks
The integers returned represent the sum of the bit weighting of the operational sta-
tus bits. These status bits are latched when the condition is detected. This register is
cleared when it is read. Refer to section 5.2.6.2 for a list of operational status bits.
RATE
Output Current Ramp Rate Setting Command
Input
RATE <rate>,[term]
Format
+n.nnnn
<rate>
Specifies the rate at which the current will ramp at when a
new output current setting is entered: 0.0001 A/s through
50.000 A/s.
Remarks
Sets the output current ramp rate. This value will be used in both the positive and
negative directions. Setting value is limited by LIMIT.
RATE?
Output Current Ramp Rate Setting Query
Input
RATE?[term]
Returned
<rate>[term]
Format
+n.nnnn (refer to command for description)