CHAPTER 4:
Operation
Model 648 Electromagnet Power Supply
4.18 Computer
Interface
There are two computer interfaces on the Model 648, a USB interface and an IEEE-
488 interface. These interfaces are used to connect the instrument to a computer for
automated control or data taking. Refer to Chapter 5. Section 4.18.1 explains how to
change the interface parameters on the IEEE-488 interface.
4.18.1 Changing IEEE-
488 Interface
Parameters
Two interface parameters, address and terminators, must be set from the front panel
before IEEE-488 communication with the instrument can be established. Other
interface parameters can be set via the interface using the device specific commands
provided in section 5.4.
Menu Navigation:
Computer Interface
Q
IEEE Address (
1 to 30
)
ENTER
Q
IEEE Term (
Cr Lf, EOI, LF, LFCR
)
Interface Command:
IEEE
4.19 Default
Parameter Values
It is sometimes desirable to reset instrument parameters to their default values. This
data is stored in nonvolatile memory called EEPROM. Instrument calibration is not
affected by this operation. Firmware version information for the main firmware and
the DAC firmware is also displayed during this sequence.
To clear EEPROM memory or view the firmware versions, press and hold
Escape
for
5 s. A screen appears to show the main firmware version, the DAC processor firmware
version, and is a prompt for returning the instrument parameters to default values.
Default parameter values are listed in TABLE 4-4.
Default parameter values
Output settings
Output current*
0 A
Current ramp rate
50.000 A/s
Maximum settings
Max output current
135.1 A
Max ramp rate
50.000 A/s
External program mode
External program mode
Internal
Ramp segments
Ramp segments
Disabled
Ramp segments current
0 A
Ramp segments rate
1.0000 A/s
Display
Brightness
25%
Keypad locking
State
Unlocked
Lock code
123
Computer interface
Baud
57,600
IEEE-488 address
12
IEEE-488 terminators
CR/LF
Mode*
Local
Water settings
Magnet water
Disabled
Internal water
Disabled
* Indicates value is also initialized on power up
Menu Navigation:
Escape
(Press and hold for 5 s)
Q
(
Yes, No
)
Interface Command:
DFLT
TABLE 4-4