5.4 Command Summary
65
|
www.lakeshore.com
LIMIT
Limit Output Settings Command
Input
LIMIT <current>,<rate>[term]
Format
+nnn.nnn,+nn.nnnn
<current>
Specifies the maximum output current setting allowed: 0 A
through 135.100 A.
<rate>
Specifies the maximum output current ramp rate setting
allowed: 0.0001 A/s through 50.000 A/s.
Remarks
Sets the upper setting limits for output current, compliance voltage, and output cur-
rent ramp rate. This is a software limit that will limit the settings to these maximum
values.
LIMIT?
Limit Output Settings Query
Input
LIMIT?[term]
Returned
<current>, <rate>[term]
Format
+nnn.nnn,+nn.nnnn (refer to command for description)
LOCK
Keyboard Lock Command
Input
LOCK <state>,<code>[term]
Format
n,nnn
<state>
0 = Unlock, 1 = Lock All, 2 = Lock Limits
<code>
Specifies lock-out code. Valid entries are 000 –999.
Remarks
Locks out all front panel entry operations.
Example
LOCK 1,123[term]
—enables keypad lock and sets the code to 123.
LOCK?
Keyboard Lock Query
Input
LOCK?[term]
Returned
<state>,<code>[term]
Format
n,nnn (refer to command for description)
MAGWTR
Magnet Water Mode Command
Input
MAGWTR<mode>[term]
Format
n
<mode>
0 = Manual Off, 1 = Manual On, 2 = Auto, 3 = Disabled
Example
MAGWTR 2[term]
—places the magnet water mode to Auto, which will automati-
cally control the magnet water valve based on the calculated output power.
MAGWTR?
Magnet Water Mode Query
Input
MAGWTR?[term]
Returned
<mode>[term]
Format
n (refer to command for description)
MODE
IEEE Interface Mode Command
Input
MODE <mode>[term]
Format
n
<mode>
0 = local, 1 = remote, 2 = remote with local lockout.
Example
MODE 2[term]
—places the Model 648 into remote mode with local lockout.