Magnet connector details
Pin
Name
1
Flow switch (com)
2
Flow switch
3
Magnet water valve A
4
Magnet water valve B
6.11.3 Auxiliary Connector
77
|
www.lakeshore.com
6.11.3 Auxiliary
Connector
The auxiliary connector provides connections for three functions:
D
Emergency stop: this normally-closed circuit turns off power to the Model 648
just as if the Off (O) button was pressed on the front panel when opened. Normal
operation requires a closed connection between the pins.
D
Fault relay: the fault relay can be used to communicate the presence of a
Model 648 fault to external equipment. The relay follows the operation of the
fault light on the front panel. Both normally-open and normally-closed configu-
rations are provided. The fault contact name represents the state of the contact
in a powered off/fault condition. For example, the normally closed (NC) fault con-
tact will be closed during a power off/fault condition, and open during normal
operation. The contacts are electrically isolated from the Model 648 chassis.
D
Remote enable: these contacts are similar in function to the flow switch inputs.
Normal (enabled) operation requires a closed connection between the pins.
Auxiliary connector
Auxiliary connector details
Pin
Name
1
Emergency stop A
2
Emergency stop B
3
Chassis–common
4
Fault–NO
5
Fault–COM
6
Fault–NC
7
Chassis–common
TABLE 6-5
FIGURE 6-8
TABLE 6-6