INSTALLATION INSTRUCTIONS
32
September 1995 – 981088
Section 3: Changing the PROCESSOR to 50 or 60 Hz Operation, M43 and
M43A Only
Important
• Clinic 1 PROCESSORS cannot be installed in 50 Hz sites. There is no 50 Hz SPROCKET for the Clinic 1
PROCESSOR. If you are installing a Clinic 1 PROCESSOR, advance to the “PROCESSOR Setup Options”
section on Page 35.
• All Kodak X-Omat M43A PROCESSORS and Kodak X-Omat Clinic 1 PROCESSORS are set up for 115 V AC,
60 Hz operation only.
• If you are installing an M43A PROCESSOR into a 50 Hz site, you will need to switch the SPROCKET on the
MAIN DRIVE MOTOR, and switch the input CONNECTOR on the DRIVE MOTOR.
• All Kodak X-Omat M43 PROCESSORS are set up for 230 V AC, 50 Hz operation.
• If you are installing an M43 PROCESSOR into a 60 Hz site, you will need to switch the SPROCKET on the
MAIN DRIVE MOTOR, and switch the input CONNECTOR on the DRIVE MOTOR.
Figure 27
Identifying the 2 SPROCKETS
H130_0167HA
CONNECTORS
50 Hz SPROCKET
20-TOOTH
60 Hz SPROCKET
16-TOOTH
H130_0167HCA