![Kodak HEALTH SCIENCES X-Omat Clinic 1 Скачать руководство пользователя страница 18](http://html.mh-extra.com/html/kodak/health-sciences-x-omat-clinic-1/health-sciences-x-omat-clinic-1_installation-instructions-manual_1984140018.webp)
INSTALLATION INSTRUCTIONS
18
September 1995 – 981088
[14] Rotate these 2 LEVELING FEET clockwise
until they are fully extended. Use a SOCKET or WRENCH to
ensure that the LEVELING FEET are tight.
[15] Position a SPACER, provided with the MOUNTING STAND, on the TOP of the MOUNTING STAND under
each of the 2 LEVELING FEET.
[16] Lower the raised end of the PROCESSOR by removing the piece of wood supporting it.
[17] Check that the LEVELING FEET are lowered into the SPACERS.
[18] Align the 2 LEVELING FEET and SPACERS with the holes in the TOP of the MOUNTING STAND. Position
the LEVELING FEET through the holes.
Figure 12
Installing the SPACERS and Replacing one LEVELING FOOT on the Non-Feed End of the
PROCESSOR
[19] Raise and support the feed end of the PROCESSOR to gain access to the other 2 LEVELING FEET. Use a 5
x 10 cm (2 x 4 in.) piece of wood that is 1 m (3 ft.) long to support the PROCESSOR.
(a) Position the piece of wood so that:
• it supports the entire width of the PROCESSOR (drive side to non-drive side)
• it is closer to the center of the PROCESSOR than to the front
[20] Remove the NUT from each of the 2 feed-end LEVELING FEET. Keep the NUTS for later installation.
[21] Rotate these 2 LEVELING FEET clockwise
until they are fully extended. Use a SOCKET or WRENCH to
ensure that the LEVELING FEET are tight.
[22] Insert the HOSES listed below through the holes in the TOP of the MOUNTING STAND:
• 3 DRAIN HOSES
- Use black, opaque DRAIN HOSES if they are going through the wall
- Use clear DRAIN HOSES if draining outside of the darkroom
• a HOSE from the main water supply, not provided
H130_0182HA
LEVELING
FOOT (2)
wood
support
H130_0182HCA
drive side
end
non-feed
SPACER (2)
MOUNTING STAND
REPLENISHMENT PUMP