Heinrichs Messtechnik GmbH
OPERATING MANUAL TMU-W / UMC4
Page 89 of 118
7.2.11.3
Measured value simulation
If the operator selected the setting “QM, D, T” described in Section 7.2.11.2 on page 88, the following three
possible settings will affect the output behaviour during measured value simulation, where all measured values
are simulated at the same time.
7.2.11.3.1
Simulation mass flow QM abs
In order to simulate mass flow, the operator can define a “measured value.” The flow rates will be simulated in
both directions. All outputs will perform based on the simulated measured value.
Set QM abs
±0900.0 kg/h
The current value will be displayed and can be changed by toggling the arrow keys.
Pressing the “Enter” key confirms the entry.
7.2.11.3.2
Density simulation
In order to simulate density/volume measurement, the operator can define a “density measured value.” If
volume measurement is assigned to an output, it will change depending on mass flow and density simulation.
All outputs will perform based on the simulated measured value.
Set density
0500.0 g/l
The current value will be displayed and can be changed by toggling the arrow keys.
Pressing the “Enter” key confirms the entry.
7.2.11.3.3
Temperature measurement simulation
In order to simulate a temperature, the operator can define a “measured value.” All outputs will perform based
on the simulated measured value.
Set temperature
+090 °C
The current value will be displayed and can be changed by toggling the arrow keys
Pressing the “Enter” key confirms the entry.