OPERATING MANUAL TMU-W / UMC4
7.2.9.2
Status output 1 assignment
This function allows the operator to define to which event the status output is to be assigned. The most general
assignment is the alarm assignment because all set limiting values and the self-test function are then
monitored via the status output.
After selecting the Status output assignment function, press “Enter” to display the current assignment.
Output 1 assigned to
[Alarm]
The current assignment will be displayed and can be changed by toggling the arrow keys.
The operator can choose between the following settings:
Flow direction recognition
o
Forward flow
o
Reverse flow
Limiting values:
o
MIN QM
o
MAX QM
o
MIN density
o
MAX density
All limiting values and error detection
o
Alarm
Pulse output 2 for custody transfer operations
o
IMP2 90°,
Pressing the “Enter” key confirms the entry.
When selecting the IMP2 90° setting, a second pulse output will be realised via the status output that can be
used for custody transfer operations.