Heinrichs Messtechnik GmbH
OPERATING MANUAL TMU-W / UMC4
Page 79 of 118
STATUS functional class
The functional class STATUS is comprised of the functions for setting the status output.
STATUS
Status output
active state
Status output 1
assignment
STATUS
functional class
Output active
[closed]
_________________
closed
open
Output 1 assigned to
[Alarm]
________________
Forw. flow
Rev. flow
MIN QM
MAX QM
MIN density
MAX density
Alarm
IMP2 90°
7.2.9.1
Status output active state
The status output can be compared to an electrical relay that can function as make or break contact. For safety-
relevant applications, the operator will choose the break contact setting so that a power failure or failure of the
electronics can be detected like an alarm. In standard applications, the output is used as make contact.
The Status output state active state function allows the operator to define the behaviour of the status output.
Status output active
[closed]
The current active state will be displayed and can be changed by toggling the arrow keys.
The operator can choose between the following settings:
closed
open
Pressing the “Enter” key confirms the entry.