Heinrichs Messtechnik GmbH
Page 104 of 118
OPERATING MANUAL TMU-W / UMC4
7.2.15.13
Pulse outputs at alarm
This parameter selects the behaviour of the pulse outputs in the event of an alarm during a flow delivery.
The following choices are available for selection:
No Alarm - Both outputs continue pulsing
Stop IMP2 - Pulse 2 is stopped. Pulse output 2 is disabled while pulse output 1 continues delivering
pulses.
Stop IMP1,2 - Both outputs are stopped, no pulses are delivered.
The default value for “pulse outputs at alarm” is [Stop IMP2]
At Alarm
[Stop IMP2]
Figure 4: Pulse behaviour in the event of an alarm
Figure 4 depicts the behaviour of the two pulse outputs in the event of an alarm during a flow delivery