KNAUER
P 6.1L User Manual V6890 Version 1.2
54
Maintenance and Care
Next steps
Insert the new inline filter cartridge.
Inserting the new Inline Filter Cartridge.
The flow direction is designated on inline filter cartridges. The
inline filter and fitting are inserted into the pressure sensor in
such a way that the designating notch always points upwards.
Prerequisite
None
Tool
Torque wrench
CAUTION!
Damage to components caused by overtightened
fittings!
Note the torque of the fittings:
- For stainless-steel fittings: 5 Nm
- For PEEK fittings: 1 Nm
Next steps
Reinstall the capillary at the pressure sensor below the fitting of
the inline filter.
3. Loosen the fitting of
the inline filter
1
with
the open-end wrench
(size 13) and remove
manually.
4. Remove clogged inline
filter cartridge.
Fig. 39
Loosen the inline filter
below the pressure sen-
sor.
1
Procedure
Process
Figure
1. Insert the inline filter
cartridge with the des-
ignating notch point-
ing upwards
1
into
the fitting
2
.
2. Manually, screw the fit-
ting with the inline fil-
ter cartridge in the
pressure sensor and
tighten with a torque
wrench using the
appropriate torque.
Fig. 40
Inserting the inline fil-
ter cartridge into the fit-
ting
1
2