P 6.1L User Manual V6890 Version 1.2
23
Installation
Connecting the Pump Head to the Solvent
Prerequisites
The device has been switched off.
The power plug has been pulled.
The front cover has been removed.
Material
Flangeless fitting
CAUTION!
Damage to the pump head possible!
Remove the cap fittings from the inlet and outlet
of the pump head prior to use!
Next steps
Integrate the pump into the HPLC flow system.
Connecting the 4-Channel Degasser to the Sol-
vent
The 4-channel degasser contains four degassing chambers. Each
degassing chamber has an inlet and an outlet on the front of the
pump.
Prerequisite
The device has been switched off.
The power plug has been pulled.
The front cover has been removed.
Material
Flangeless fitting
CAUTION!
Very high pressures damage the degasser mem-
brane. The membrane can withstand a maximum
pressure of 7 bar.
Do not connect the degasser to the pump outlet.
Procedure
Process
Figure
1. Slide the flangeless fit-
ting onto the tubing.
2. Insert the tubing into
the free inlet
1
on the
bottom of the pump
head.
3. Tighten the fitting by
hand.
Fig. 15
Solvent line on the pump
head
1