KNAUER
P 6.1L User Manual V6890 Version 1.2
57
Technical Data
P 6.1L Quaternary
Gradient range
0–100 %
5–95 % (recommended)
Minimum incre-
ment
0.1 %
HPG: gradient
accuracy
± 0.3 % (measured at 1 ml/min, 150 bar,
ethanol/caffeine tracer)
± 1 % (5–95 %, measured at 0.1–10 ml/
min, water/caffeine tracer)
HPG: gradient
precision
< 0.1 % RSD (measured at 1 ml/min,
0.3 % RSD overall, based on retention
time at constant room temperature)
Mixer
Mixing volume
50, 100, 200 μl
250 μl (metal-free)
Delay volume
110 μl (dependent of mixer)
410 μl (metall-free)
Setup
Pump type
Quaternary analytical HPLC pump with
degasser
Variants of the
pump head
5 ml/min stainless steel
10 ml/min stainless steel
10 ml/min ceramic
Degasser
4 channels, Teflon® AF
Special feature
Automatic adaption of LPG cycle time
Weight
Weight
12.7 kg
Gradient formation
Gradient forma-
tion
Low-pressure quaternary mixing
Gradient range
0–100 %
1–99 % (recommended)
Minimum
increment
0.1 %.
LPG: gradient
accuracy
± 0.3 % (measured at 1 ml/min,
150 bar, ethanol/caffeine tracer)
± 2 % (1–99 %, measured at 5–50 % of
the flow range, water/caffeine tracer)
LPG: gradient
precision
< 0.1 % RSD (measured at 1 ml/min,
0.5 % RSD overall, based on retention
time at constant room temperature)
Mixer
Mixing volume
50, 100, 200 μl
250 μl (metal-free)
Delay volume
210 μl (dependent of mixer)
410 μl (metal-free)