KNAUER
P 6.1L User Manual V6890 Version 1.2
31
Installation
Events Connector
For test purposes or in some other cases, it can make sense to
manually enter these signals:
Sending control signals (Events) to external devices
Opening and closing contacts
Activating 500 ms pulses
Error OUT
TTL output
Levels:
passive 5 V
active 0 V
Error IN
TTL input
Low active
Secure switching threshold at least 10 mA
After receiving a signal (short-circuit to
ground) from an external device, an error
message appears and the device stops.
Error GND
Reference point of the voltage at the signal
inputs.
+24V
Event-controlled switching of 24 V
against GND
Protection: 24 V – 200 mA
+5V
Provides a voltage of 5 V with respect to
GND. This makes it possible to supply a
consumer that is switched by an EVENT.
Protection: 5 V – 50 mA
GND
Reference point of the voltage at the signal
inputs.
Signal
Explanation
Legend
1
Display
2
Events
Remote Con-
nector
Fig. 28
Relation between display and
Events
terminal strip
1
2