KNAUER
P 6.1L User Manual V6890 Version 1.2
55
Maintenance and Care
Replacing the Mixer
A clogged mixer can cause pressure fluctuations and irregular
flow. The mixer is replaced completely as an assembly.
Procedure
1. Remove capillary from the inlet and outlet of the mixer.
2. Remove the screws from the front of the mixer.
3. Remove the mixer
4. Install new mixer
Prerequisite
None
Tools
Open-end wrench, size 1/4"
Allen wrench, 2 mm and 2.5 mm
Torque wrench
CAUTION!
Damage to components caused by overtightened
fittings!
Note the torque of the fittings:
- For stainless-steel fittings: 5 Nm
- For PEEK fittings: 1 Nm
Next steps
Bring the device into operation again.
Procedure
Process
Figure
1. Remove capillary
2
at
the inlet and outlet of
the mixer
2. Remove the screws
1
from the front of the
mixer and remove the
mixer.
3. Install new mixer.
4. Manually, screw the fit-
tings of the capillary in
the mixer.T
5. ighten with a torque
wrench using the
appropriate torque.
Fig. 41
Mixer
Fig. 42
Mixer bio
1
2
1
2