KNAUER
P 6.1L User Manual V6890 Version 1.2
45
Troubleshooting
"Pump head type: write
failed"
Restart the device.
Repeat the automatic config-
uration step in the chroma-
tography software.
Remove pump head, clean it
and re-install it.
If this system message occurs
repeatedly, contact the Tech-
nical Support.
S
"SetPoint in the program is
not compatible with the
current PH PMax"
Note the maximum pressure of
the pump head.
T
"This link is used in
WAKEUP"
First quit or delete wakeup pro-
gram (wu = Wake Up), then edit
or delete link.
"This program is used in a
link"
First pause or delete the link,
then edit or delete data using
the chromatography software.
"This program is used in
WAKEUP"
First quit or delete wakeup pro-
gram (wu = Wake Up), then edit
or delete data by means of the
chromatography software.
"Too many lines in
program"
Check the number of program
lines.
A maximum of 100 program
lines are possible.
U
"Unable to attain min.
flow setpoint"
Confirm, pump continues run-
ning.
"Unable to attain pressure
setpoint"
The entered pressure cannot be
achieved with the maximum
flow set in the
Constant Pressure
mode.
Check for leaks.
Increase the upper
flow level.
Reduce the working pressure.
"Unknown pump head
type"
Check the pump head.
Check whether a pump head
with RFID recognition has
been installed.
If this system message occurs
repeatedly, contact the Tech-
nical Support.
W
"Wake up time already
passed"
Correct the entry for date or
otherwise time.
System message