Installation
Note
To integrate the pump into a system, note the ambient condi-
tions found in the sections Operating Environment and Techni-
cal Data as well as the ambient conditions of other devices to be
integrated into that system.
CAUTION!
Damage to the pump head caused by overtight-
ened capillary fittings!
Note the torque of the fittings:
- 5 Nm for stainless-steel fittings
- 0.5 Nm for PEEK fittings
Tool
Torque wrench
Procedure
Stainless steel fittings are tightened with 5 Nm, PEEK fittings are
tightened with 0.5 Nm.
Isocratic Pump
The figure shows the installation plan for capillary and tubing on
the isocratic version of the pump. If you own a different version
of the pump, see the corresponding chapter for installation
instructions.
The gray lines represent the connection of the solvents to the
pump and the black lines represent the connection of the piston
backflushing.
Piston backflushing
The flushing solution is re-used. Since the flow path is circular,
only one bottle is used for the flushing solution.
Solvent flow path
The pump head takes the liquid in from the bottle and conveys
it to the pressure sensor.
Connecting the Pump Head to the Solvent
Prerequisites
The device has been switched off.
The power plug has been pulled.
The front cover has been removed.
Material
Flangeless fitting
Capillary layout
Fig. 9
Installation plan for the isocratic pump