KNAUER
P 6.1L User Manual V6890 Version 1.2
18
Installation
Pre-Installed Capillary
Pre-installed stainless steel and PEEK capillary is color-coded
according to its inner diameter.
Note
PEEK capillary is not suitable for use with Acetonitrile. Acetoni-
trile can cause capillary to crack or rupture.
Integrating the Pump into an HPLC Flow Sys-
tem
The pump can be integrated into an HPLC flow system by con-
necting the pressure sensor (isocratic version) or the mixer
(binary or quaternary version) and the HPLC system with capil-
lary.
Procedure
Process
Figure
1. Slide the flangeless fit-
ting
1
over the tub-
ing.
2. Slide the lock ring
2
over the tubing, cap
with the sealing ring
3
. Note the direction
of the lock ring, other-
wise the sealing ring
could be damaged.
The thicker end of the
lock ring must point
into the direction of
the fitting. Fasten the
polymer seal.
Fig. 7
Set-up of a flangeless fit-
ting
3. Manually fasten the
flangeless fitting
4
to
the device.
Fig. 8
Fully assembled flange-
less fitting
1
2
3
4
Color
Material
Inner diameter
red marker
Stainless steel
0.1 mm
blue marker
Stainless steel
0.25 mm
black marker
Stainless steel
0.45 mm
blue stripes
PEEK
0.25 mm
orange stripes
PEEK
0.5 mm