Troubleshooting
"Not enough space to
store link"
Check the pump.
Check the number of pro-
gram lines - a maximum of
100 program lines are possi-
ble.
"Not enough space to
store program"
Check the pump
Check the number of pro-
gram lines - a maximum of
100 program lines are possi-
ble.
"Not in HPG mode"
Select HPG mode.
P
"Pressure max in the pro-
gram is not compatible
with the current pump
head"
Note the maximum pressure of
the pump head.
"Program does not exist,
please edit the program
number"
Create and edit a program.
"Program is running"
Quit program or wait until pro-
gram has been completed.
"Pump head type: head
data uninitialized"
Restart the device.
Check whether a pump head
with RFID recognition has
been installed.
Repeat the automatic config-
uration step in the chroma-
tography software.
Remove pump head, clean it
and re-install it.
"Pump head type:
read failed"
Restart the device.
Repeat the automatic config-
uration step in the chroma-
tography software.
Remove pump head, clean it
and re-install it.
If this system message occurs
repeatedly, contact the Tech-
nical Support.
"Pump head type: RFID
hardware not present or
failed"
Pump head without RFID detec-
tion: If necessary, replace pump
head.
System message