Safety for Users
If you do not belong to this or a comparable professional group,
you may not perform the work described in this user manual
under any circumstances. In this case, please contact your supe-
rior.
Safety Equipment
When working with the device, take measures according to lab
regulations and wear protective clothing:
Safety glasses with side protection
Protective gloves
Lab coat
What must be taken into
account?
All safety instructions in the user manual
The environmental, installation, and connection specifica-
tions in the user manual
National and international regulations pertaining to labora-
tory work
Original spare parts, tools, and solvents made or recom-
mended by KNAUER
Good Laboratory Practice (GLP)
Accident prevention regulations published by the accident
insurance companies for laboratory work
Filtration of substances under analysis
Use of inline filters
Once they have been used, never re-use capillaries in other
areas of the HPLC system.
Only use a given PEEK fitting for one specific port and never
re-use it for other ports. Always install new PEEK fittings on
each separate port.
Follow KNAUER or manufacturer's instructions on caring for
the columns
More safety-relevant information is listed in alphabetical order in
the following table:
Topic
Explanations
flammability
Organic solvents are highly flammable.
Since capillaries can detach from their
screw fittings and allow solvent to escape,
it is prohibited to have any open flames
near the analytical system.
solvent tray
Risk of electrical shock or short circuit if
liquids get into the device's interior. For
this reason, place all bottles in a solvent
tray.
solvent lines
Install capillaries and tubing in such a way
that liquids cannot get into the interior in
case of a leak.
leaks
Regularly check if any system compo-
nents are leaking.