Suspension Arm Systems Independant
Revision 1
111
9.2.4
General Technical Data
Designation
All suspension arm systems
Operating mode
suitable for continuous operation
Activation time of the electromagnetic brakes
max. 1 min*
Rating plate
The rating plate is located on the upper extension
arm, see section 4.1 “Data on the Rating Plate”,
page 30.
Electrical data
Nominal voltage
100–240 VAC
Nominal frequency
60/50 Hz
Rated output
max. 220 W
Indirect extension arm lighting
2–4 light panels
(operating voltage 12 VDC, each with 2 light
panels connected in series to 24 VDC)
12 VDC
Sound pressure level
A level of 65 db(A) (EN ISO 3746) is not exceeded
Manual forces (when the electromagnetic brake
is released depending on position and payload)
3.5–40 Nm
Protection class acc. to IEC 60601-1
I
IP classification as per IEC 60529
IP 20
Approvals of the standard design
Recognized NRTL component
Ambient conditions
Storage and transportation
-25°C to +70°C (-13°F to +158°F) at 10–75%
humidity (relative) and air pressure 500–1,060 hPa
Operation
+10°C to +40°C (+50°F to +104°F) at 30–75%
humidity (relative) and air pressure 700–1,060 hPa
(corresponds to a maximum operating altitude of
3,000 m)
Conformity to standards
see section 3.5 “Standards and Guidelines”, page 28
Medical Device Directive 93/42/EWG
Classification I
Permissible adaptions
see section 9.3 “Permissible Adaptions”, page 112
*
If the electromagnetic brakes are actuated for a longer time, the power supply unit can switch off automatically to
prevent overheating.
If the power supply unit has switched off, then it must cool down for 10 min before it can be switched on again,
and must then be disconnected from the building’s power supply for 10 seconds.
Only then can the standard operation of the system be resumed.