Suspension Arm Systems Independant
60
Revision 1
WARNING
Danger of injury by trapped limbs or from collisions!
Limbs may be trapped during assembly work. Components of the suspension arm system
may cause collisions.
•
During the installation, be particularly careful.
WARNING
Danger of injury from falling parts!
During the installation, parts may fall.
During the installation, no-one must stand beneath the components of the suspension arm
system.
•
Securely position the flange of the extension arm
,
on a suitable lifting device and
position under the threaded bolts M16 of the interface plate
.
NOTICE
Position of the flange for the Duo version
The flanges have a chamfered surface to allow installation on the same height.
•
Mount the flanges in a way that the chamfered surfaces point towards each other.
•
Insert flange of the ceiling tube with extension arm
,
into the 6 threaded bolts
of the interface plate
.
•
Remove the previously attached adhesive tape or rubber tape from the threaded bolts .
•
For each threaded bolt attach 1 plastic insulation washer as per the illustration in a
way that the retaining washer with external Ø 34 mm (see arrow in the enlarged detail
view of the illustration) sits in the plastic insulation washer .
•
For each threaded bolt use one 1 retaining washer , spring washer and
1 hexagonal nut M16 .
Components on the interface plate
:
•
Adjust the Single version, see section 7.3.3 “Adjusting the Extension Arm on the Single
Interface Plate”, page 61.
•
Adjust the Duo version, see section 7.3.4 “Adjusting the Extension Arm on the Duo
Interface Plate”, page 63.