114
Revision 1
Not OK OK
•
Gas modules (if provided) are firmly installed and functional**
•
Have gas modules and supply lines (if provided) inspected by an authorized
entity for gas type and gas density**
•
The installed hoses and supply lines must be inspected annually for visible
damage. Replacement is required if visible damage is present, or no later
than after 10 years.
•
Brake guides are lit (optionally) when the corresponding brake button is
pressed*
•
Suspension arm lighting (optional) lights up when the switch is pressed (on
the M6 service head or on the CEMOR
®
)**
additionally for IDP 101/201 M
Not OK OK
•
The motorized height adjustment moves smoothly along the selected travel
path**
•
The upper stop functions and is set correctly, readjust if necessary**
•
The lower stop functions and is set correctly, readjust if necessary**
•
The trapezoidal spindle runs smoothly and is adequately greased;
regrease with Gleitmo 810 if required*
additionally for IDP 101 S
Not OK OK
•
Load compensation/spring force correct, readjust as needed**
Comments:
The work listed above was performed, including the required adjustment work and safety
inspection:
Date
Name (in block letters)
Signature/Stamp
* Any damaged, deformed or missing components should be replaced by way of precaution.
Please contact the suspension arm system supplier for this.
** If any of the items should be objected to during the inspection, then the suspension arm system must
be shut down by way of precaution to rule out further personal injury or equipment damage. Contact
the system supplier immediately.
*** Does not apply to systems with electromagnetic brakes
The medical device book accompanying every medical device and mandated by the German
Medical Products Operator Ordinance (MPBetreibV) must be kept on-site. Service and
maintenance work as well as safety inspections must be recorded in this medical device book.
Test reports such as this one must be archived in the respective medical device book.