Suspension Arm Systems Independant
Revision 1
67
Use fastening screws suitable for the type of ceiling tube
IDP 401 ceiling tube:
•
8 Allen countersunk screws M10 x 25 mm (10.9) DIN EN ISO 10642.
IDP 501 XL ceiling tube:
•
8 Allen countersunk screws M10 x 25 mm (10.9) DIN EN ISO 10642
(IDP 501 XL, Single arm)
•
8 Allen countersunk screws M10 x 30 mm (10.9) DIN EN ISO 10642
(IDP 501 XL, Dual arm)
IDP 501 XXL ceiling tube:
•
8 Allen countersunk screws M10 x 30 mm (10.9) DIN EN ISO 10642
•
Screw the console tube
through the fastening plate on the extension arm
,
using 8 countersunk screws .
WARNING
Danger of injury due to the suspension arm system falling!
Insufficiently tightened fastening elements can cause the suspension arm system to fall:
•
Tighten 8 countersunk screws with a torque of 40 Nm.
Make sure the console tube
is securely in place:
•
The console tube
must sit flush on the extension arm
,
.
•
The countersunk screws must be tightened with 40 Nm.