126
TOS3200
Troubleshooting
Unable to make correct measurements.
Unable to control the TOS3200 remotely through the RS232C.
*1
See the Communication Interface Manual.
Symptom
Check and Remedy
Page
The measured values are not
correct.
• Is the test lead connected to the correct terminal?
Connect a test lead to the measurement terminal for which the LED is
illuminated.
• Is the measurement network that you are using correct?
Use the measurement network required by the standard to make the
measurement.
–
• Is the CONV function turned on?
Turn CONV off to determine the measured values at the power
voltage supplied to the EUT.
• Is WAIT turned OFF?
If you do not measure the EUT data at startup, turn WAIT on and set
an appropriate time.
• If the power cord of the EUT is a two-prong cord with a ground wire,
is the ground line connected to a ground terminal?
• Is the EUT connected to both the plug on the front panel and the
terminal block on the rear panel?
–
The measured values are not
updated.
• Is the TOS3200 set to hold the measured value?
If MEAS MODE is set to MAX in Meter Mode, press the MAXCLR
(F4) key to clear the held, measured value.
Unable to save the test data
when the judgment is PASS.
• Is the PASS hold time too short?
The test does not stop.
• Is the timer turned off?
The specified test time is not enabled. Turn the timer on.
–
The program test does not
stop.
• Is there a step in which the timer is turned off?
The program will advance to the next step only if U-FAIL or
CONTACT FAIL occurs for steps whose timer is turned off. Turn the
timer on.
–
The test does not stop even if
the judgment is FAIL in the
middle of a program test.
• Is ABORT turned off?
Turn ABORT on if you want to stop the test on a FAIL judgment at the
step level.
The time when the test data
was saved is not correct.
• Is the system clock correct?
Set the system clock to the present time.
See
Symptom
Check and Remedy
Page
Unable to control the
TOS3200 from the PC.
• Is the talk mode turned on?
Turn talk mode off to control the TOS3200 from a PC.
*1
• Are you using a cross cable?
*1
See
Содержание TOS3200
Страница 15: ...General Description This chapter gives an overview and describes the features...
Страница 20: ...20 TOS3200 Options This page has been intentionally left blank...
Страница 29: ...Panel Control Basics This chapter explains how to select items on the screen and how to enter values...
Страница 38: ...38 TOS3200 Panel Memory This page has been intentionally left blank...
Страница 67: ...Meter Mode Measurement This chapter explains the touch current measurement in Meter Mode Meter Mode Measurement...
Страница 73: ...Program Test This chapter describes how to create and edit sequence pro grams and how to execute the test...
Страница 105: ...Specifications This chapter gives the specifications and external dimensions...
Страница 115: ...Appendix A A List of Default Settings B System Settings C Protection Function D Troubleshooting...
Страница 129: ......