48
TOS3200
Judgment System
The TOS3200 judges PASS, L-FAIL, or U-FAIL with respect to a preset reference.
Table 4-2 shows the judgments for measured values according to the TIMER, LOWER, and
UPPER settings. Pass and L-FAIL are judged only if TIMER is ON.
Validity of the Lower Reference
Normally, the TC and PCC of the same EUT are within a specified range of the EUT, but if the
current is extremely small, it cannot be measured. Specifying LOWER ON and setting the
lower reference to a value slightly smaller than the range specific to the EUT help to detect a
disconnection or a bad connection of the test leads. This allows more reliable tests.
Table 4-2
Judgment patterns
• PASS
If the measured values are within the reference range up to the end of the
test (TIMER 0 s), the TOS3200 will indicate PASS, and the test ends.
• L-FAIL
If a current less than or equal to the lower reference is measured, the
TOS3200 will indicate L-FAIL, but the test continues until the end of the test
(TIMER 0 s).
• U-FAIL
If a current greater than or equal to the upper reference is measured, the
TOS3200 will indicate U-FAIL, and the test stops immediately.
TIMER
LOWER
UPPER
Measured value distribution
judgment
Lower reference
Upper reference
ON
ON
ON
PASS
U-FAIL
L-FAIL
U-FAIL
OFF
ON
PASS
U-FAIL
ON
OFF
PASS
L-FAIL
OFF
OFF
PASS
Содержание TOS3200
Страница 15: ...General Description This chapter gives an overview and describes the features...
Страница 20: ...20 TOS3200 Options This page has been intentionally left blank...
Страница 29: ...Panel Control Basics This chapter explains how to select items on the screen and how to enter values...
Страница 38: ...38 TOS3200 Panel Memory This page has been intentionally left blank...
Страница 67: ...Meter Mode Measurement This chapter explains the touch current measurement in Meter Mode Meter Mode Measurement...
Страница 73: ...Program Test This chapter describes how to create and edit sequence pro grams and how to execute the test...
Страница 105: ...Specifications This chapter gives the specifications and external dimensions...
Страница 115: ...Appendix A A List of Default Settings B System Settings C Protection Function D Troubleshooting...
Страница 129: ......