84
TOS3200
Setting Additional Test Conditions
You can set the following functions in the system setup screen (SYSTEM1/5) in addition to
the program test conditions.
• Maximum value hold function of the measured current (MEAS MODE)
• Hold time of the PASS judgment (PASS HOLD)
• Measured current conversion display (CONV)
• Buzzer volume (BUZ VOL)
• Power supply to the EUT during the interval time (LINE BRK (AUTO))
Maximum Value Hold Function of the Measured Current
(MEAS MODE)
This function holds and displays the maximum value measured during the test period.
The maximum value is held for a single test. The held value is cleared when the test is com-
pleted.
The maximum measured value is not held during the test wait time.
1
Select the MEAS MODE on SYSTEM1/5.
2
Use the rotary knob to select the NORM or MAX.
Hold Time of the PASS Judgment (PASS HOLD)
The test result data can be saved to the memory when the total judgment is displayed after
the program test is completed.
If the total judgment is FAIL, the screen will show the FAIL result until you press the STOP
switch. Therefore, you can save the data during this period.
On the contrary, the display time when the total judgment is PASS is 2.0 s (default value).
Thus, you must save the test result data during this period. If you want to make sure to save
the data for PASS judgments, set the display time of the PASS judgment result (PASS HOLD)
to HOLD.
SYSTEM
MANUAL
I / F
EDIT
AUTO
MEAS
PASS
CONV
MEAS MODE :NORM
BUZ VOL (PASS)
: 3
PASS HOLD : 2.0s
BUZ VOL (FAIL)
: 3
CONV :OFF
CONTRAST
:
5
SELV :OFF LINE
BRK(AUTO)
:ON
SELV
<NEXT>
SYSTEM1 / 5
See
NORM
Normal measurement (not hold the maximum value)
MAX
Hold the maximum value.
See
See
Содержание TOS3200
Страница 15: ...General Description This chapter gives an overview and describes the features...
Страница 20: ...20 TOS3200 Options This page has been intentionally left blank...
Страница 29: ...Panel Control Basics This chapter explains how to select items on the screen and how to enter values...
Страница 38: ...38 TOS3200 Panel Memory This page has been intentionally left blank...
Страница 67: ...Meter Mode Measurement This chapter explains the touch current measurement in Meter Mode Meter Mode Measurement...
Страница 73: ...Program Test This chapter describes how to create and edit sequence pro grams and how to execute the test...
Страница 105: ...Specifications This chapter gives the specifications and external dimensions...
Страница 115: ...Appendix A A List of Default Settings B System Settings C Protection Function D Troubleshooting...
Страница 129: ......