106
TOS3200
Specifications
Unless specified otherwise, the specifications are for the following settings and conditions.
• The warm-up time is 30 minutes.
• rdng: Indicates the read value.
• EUT: Denotes the equipment under test.
Measurement Items, Measurement Mode, and Measurement Network
Current Measurement Section
*1
Measurement
item
TC
TC measurement
Measurement method
Uses a measurement network representing the human body impedance,
measures the voltage drop across the reference resistance, and calcu-
lates the TC.
PROBE
setting
E
NC
P
E
Measurement terminal A: Measurement terminal (for connecting to the
EUT enclosure)
Measurement terminal B: Open
E
NC
E
NC
Measurement terminal A and B: Measurement terminals (for connecting
to the EUT enclosure)
E
NC
L
IV
,
E
NC
N
EU
Measurement terminal A: Measurement terminal (for connecting to the
EUT enclosure)
Measurement terminal B: Open
PCC
Protective conductor current measurement
Measurement method
Measures the voltage drop across a reference resistance that is inserted
in the middle of the protective earth wire and measure the PCC.
METER
Measures the current flowing through measurement terminals A and B or
the applied voltage (cannot be measured simultaneously).
Measurement
method
Current mea-
surement
Uses a measurement network representing the human body impedance,
measures the voltage drop across the reference resistance, and calcu-
lates the current flowing through terminals A and B.
Voltage mea-
surement
Measures the voltage applied between measurement terminals A and B.
Measurement
mode
(MODE)
DC
Measures only the DC component by eliminating the AC component.
RMS
Measures the true rms value.
PEAK
Measures the waveform peak value.
Measurement
network
(NTWK)
Network A (IEC 60990 compliant)
Basic measurement element:(1.5 k
Ω
// 0.22 µF) + 500
Ω
Network B (IEC 60990 compliant)
Basic measurement element:(1.5 k
Ω
// 0.22 µF) + 500
Ω
// (10 k
Ω
+ 22
nF)
Network B1 (IEC 60990 compliant)
Network C (IEC 60990 compliant)
Basic measurement element:
(1.5 k
Ω
// 0.22 µF) + 500
Ω
// (10 k
Ω
+ (20 k
Ω
+ 6.2 nF) // 9.1 nF)
Network D
Basic measurement element:1 k
Ω
Network E
Basic measurement element:1 k
Ω
// (10 k
Ω
+ 11.225 nF + 579
Ω
)
Network F
Basic measurement element:1.5 k
Ω
// 0.15 µF
Network G
Basic measurement element:2 k
Ω
Network constant tolerance
Resistor: ±0.1 %, capacitor 0.15 µF: ±2 %, others: ±1 %
Display/
resolution
i: measure-
ment current
i < 1 mA
µA / 1 µA
1 mA
≤
i < 10 mA
.
mA / 0.01 mA
10 mA
≤
i < 100 mA
.
mA / 0.1 mA
See
Содержание TOS3200
Страница 15: ...General Description This chapter gives an overview and describes the features...
Страница 20: ...20 TOS3200 Options This page has been intentionally left blank...
Страница 29: ...Panel Control Basics This chapter explains how to select items on the screen and how to enter values...
Страница 38: ...38 TOS3200 Panel Memory This page has been intentionally left blank...
Страница 67: ...Meter Mode Measurement This chapter explains the touch current measurement in Meter Mode Meter Mode Measurement...
Страница 73: ...Program Test This chapter describes how to create and edit sequence pro grams and how to execute the test...
Страница 105: ...Specifications This chapter gives the specifications and external dimensions...
Страница 115: ...Appendix A A List of Default Settings B System Settings C Protection Function D Troubleshooting...
Страница 129: ......