TOS3200
117
A List of Default Settings
Appx
Meter mode
measurement
conditions
(METER)
NTWK
CURR:NETW
A
←
Depends on the
memory contents.
MODE
CURR:MODE
RMS
VOLT:MODE
RMS
RANGE
CURR:RANG:SEL
AUTO
System
settings
(SYSTEM)
MEAS MODE
SYST:CONF:MMOD
NORM
No change
←
PASS HOLD
SYST:CONF:PHOL
2.0 s
CONV
SYST:CONF:CONV
OFF
SELV
SYST:CONF:SELV
OFF
BUS VOL (PASS)
SYST:BEEP:VOL:PASS
3
BUS VOL(FAIL)
SYST:BEEP:VOL:FAIL
3
CONTRAST
DISP:CONT
5
LINE BRK (AUTO) SYST:CONF:LBR
ON
TIME ADJUST
SYST:DATE /
SYST:TIME
No change
*1
CAL. DATE
–
ALARM
–
CAL. PROTECT
–
ON
Interface
settings
(INTERFACE)
I/F SELECT
–
GPIB
No change
←
GPIB ADDRESS
–
3
BAUDRATE
–
19200
DATA BITS
–
8
X-FLOW
–
ON
STOP BITS
–
1
TALK MODE
–
OFF
ERR TRACE
SYST:ERR:TRAC
OFF
Individual test data
–
No saved data
No change
←
Program test data
–
No saved data
Panel memory
–
See .
Sequence program
–
No registered
steps
Trigger source
(during remote control)
TRIG:SOUR
IMM
←
←
Sets the program number
(during remote control)
PROG:NAME
-1
*2
←
←
Power supply line
OUTP:LINE
OFF
←
←
Display during testing
DISP:SIZE
NORM
No change
←
Voltage display in expanded
display mode
DISP:UXV
OFF
No change
←
*1.
The factory default settings are shown below.
TIME ADJUST:
Standard Japanese time at factory shipment.
CAL. DATE:
Calibration date at factory shipment.
ALARM:
One year after CAL. DATE.
*2.
Aborts the program operation.
Item
Command
Setting after
initialization
Setting after sending the
command
*RST
*RCL
Содержание TOS3200
Страница 15: ...General Description This chapter gives an overview and describes the features...
Страница 20: ...20 TOS3200 Options This page has been intentionally left blank...
Страница 29: ...Panel Control Basics This chapter explains how to select items on the screen and how to enter values...
Страница 38: ...38 TOS3200 Panel Memory This page has been intentionally left blank...
Страница 67: ...Meter Mode Measurement This chapter explains the touch current measurement in Meter Mode Meter Mode Measurement...
Страница 73: ...Program Test This chapter describes how to create and edit sequence pro grams and how to execute the test...
Страница 105: ...Specifications This chapter gives the specifications and external dimensions...
Страница 115: ...Appendix A A List of Default Settings B System Settings C Protection Function D Troubleshooting...
Страница 129: ......