80
TOS3200
Creating Sequence Programs
Setup Items of AUTO2/2 (Common to All Steps)
Press the NEXT (SHIFT+ ) key to display the program edit screen 2/2 (AUTO2/2).
On AUTO2/2, set the items common to all steps.
*1
Measures voltage U1 between the measurement network reference points.
Table 6-2
Determination of the fixed range
SHIFT
PREV
NEXT
AUTO 2 / 2
PRG
00:--UNTITLED--
EDIT
Rs: 1.5 k
Ω
Cs: 0.22 μF
Rb: 0.5 k
Ω
RMS
AUTO
NTWK MODE
RANGE
OFF
ABORT
TITLE
NTWK
MODE
RANGE
ABORT
A
A
B
Rs
Cs
Rb
11
12
13
15
14
Item
Description
Panel operation
11
TITLE
Selects the program name.
F1 key
Name
Up to 12 of the characters shown in Table 3-1 on page 37 can be
entered.
Rotary knob
12
NTWK
Selects the measurement network.
F2 key
A
(For IEC 60990) (1.5 k
Ω
// 0.22 µF) + 500
Ω
SHIFT+F2 key
B
(For IEC 60990) (1.5 k
Ω
// 0.22 µF) + 500
Ω
// (10 k
Ω
+ 22 nF)
B1
*1
(For IEC 60065) (1.5 k
Ω
// 0.22 µF) + 500
Ω
// (10 k
Ω
+ 22 nF)
C
(For IEC 60990)
(1.5 k
Ω
// 0.22 µF) + 500
Ω
// (10 k
Ω
+ (20 k
Ω
+ 6.2 nF) // 9.1 nF)
D
1 k
Ω
E
1 k
Ω
// (10 k
Ω
+ 11.225 nF + 579
Ω
)
F
1.5 k
Ω
//0.15 µF
G
2 k
Ω
13
MODE
Selects the current measurement mode.
F3 key
RMS
RMS measurement
SHIFT+F3 key
DC
DC measurement
PEAK
Peak measurement
14
RANGE
Selects the measurement range.
F4 key
AUTO
Automatically switches the range according to the measured
value.
SHIFT+F4 key
FIX
Fixes the range. The measurement range is determined by the
upper reference, measurement network (NTWK), and current
measurement mode (MODE) settings. See Table 6-2.
15
ABORT
Selects the operation if a FAIL judgment occurs while a program test is in
progress.
F5 key
OFF
Executes the next step even if a FAIL judgment occurs.
SHIFT+F5 key
ON
Aborts the program test when a FAIL judgment occurs.
See
See
TC measurement
PCC measurement
Measurement
range
MODE
Upper reference
Upper reference
NTWK A, B, B1, C
NTWK D, E
NTWK F
NTWK G
Range 1
DC, RMS
30 µA to 600 µA
30 µA to 300 µA
30 µA to 200 µA
30 µA to 150 µA
30 µA to 600 µA
PEAK
50 µA to 850 µA
50 µA to 424 µA
50 µA to 282 µA
50 µA to 212 µA
50 µA to 850 µA
Range 2
DC, RMS 601 µA to 6.00 mA
301 µA to 3.00 mA 201 µA to 2.00 mA 151 µA to 1.50 mA 601 µA to 6.00 mA
PEAK
851 µA to 8.50 mA
425 µA to 4.24 mA 283 µA to 2.82 mA 213 µA to 2.12 mA 851 µA to 8.50 mA
Range 3
DC, RMS 6.01 mA to 30.0 mA 3.01 mA to 30.0 mA 2.01 mA to 20.0 mA 1.51 mA to 15.0 mA 6.01 mA to 30.0 mA
PEAK
8.51 mA to 90.0 mA 4.25 mA to 45.0 mA 2.83 mA to 30.0 mA 2.13 mA to 22.5 mA 8.51 mA to 90.0 mA
Содержание TOS3200
Страница 15: ...General Description This chapter gives an overview and describes the features...
Страница 20: ...20 TOS3200 Options This page has been intentionally left blank...
Страница 29: ...Panel Control Basics This chapter explains how to select items on the screen and how to enter values...
Страница 38: ...38 TOS3200 Panel Memory This page has been intentionally left blank...
Страница 67: ...Meter Mode Measurement This chapter explains the touch current measurement in Meter Mode Meter Mode Measurement...
Страница 73: ...Program Test This chapter describes how to create and edit sequence pro grams and how to execute the test...
Страница 105: ...Specifications This chapter gives the specifications and external dimensions...
Страница 115: ...Appendix A A List of Default Settings B System Settings C Protection Function D Troubleshooting...
Страница 129: ......