TOS3200
123
System Settings
Appx
Setup Items of SYSTEM5/5
On SYSTEM4/5, press the <NEXT> (F5) key to display SYSTEM5/5.
On SYSTEM5/5, set the items indicated in below.
Item
Description
Panel operation
3
DELETE
Deletes the contents of the specified test data.
Displayed only when a memory number containing test data is speci-
fied.
F3 key
4
<PREV>
Displays SYSTEM3/5.
F4 key
5
<NEXT>
Displays SYSTEM5/5.
F5 key
ADJUST
ALARM
TIME ADJUST :2006/10/18 16:26
CAL. DATE
:2006/10/07
ALARM :2007/10/18
CAL.PROTECT :ON
<PREV>
SYSTEM5 / 5
1
2
3
4
5
Item
Description
Panel operation
1
TIME ADJUST
Sets the system clock.
Enter the present time, and press the ADJUST (F1) key to apply the
entered time.
F1 key
Time
Sets the present time.
Year/month/day hour:minute
Rotary knob
2
CAL. DATE
Set to the date when calibration service was performed. (You cannot
set this date.)
Year/month/day
3
ALARM
Sets the calibration time limit.
F3 key
Time
Sets the next calibration date.
Year/month/day
Rotary knob
4
CAL.
PROTECT
Sets the action taken when the calibration time limit expires.
ON
The message “CAL PROTECTION” will blink when the
power is turned on if the calibration date is due.
You will not be able to use the TOS3200 if “CAL PROTEC-
TION” is displayed. Press the STOP switch to display
SYSTEM5/5, and turn CAL. PROTECT off to continue
using the TOS3200.
SHIFT+F4 key
OFF
The message “CAL DATE EXPIRED” will appear when the
power is turned on if the calibration date is due.
Press the STOP switch to continue using the TOS3200.
5
<PREV>
Displays SYSTEM4/5.
F4 key
Содержание TOS3200
Страница 15: ...General Description This chapter gives an overview and describes the features...
Страница 20: ...20 TOS3200 Options This page has been intentionally left blank...
Страница 29: ...Panel Control Basics This chapter explains how to select items on the screen and how to enter values...
Страница 38: ...38 TOS3200 Panel Memory This page has been intentionally left blank...
Страница 67: ...Meter Mode Measurement This chapter explains the touch current measurement in Meter Mode Meter Mode Measurement...
Страница 73: ...Program Test This chapter describes how to create and edit sequence pro grams and how to execute the test...
Страница 105: ...Specifications This chapter gives the specifications and external dimensions...
Страница 115: ...Appendix A A List of Default Settings B System Settings C Protection Function D Troubleshooting...
Страница 129: ......