90
TOS3200
Stopping the Test
Program Test Judgment (Cont’d)
Judgment for each step
Judgment is performed in the same manner as the single test. For details, see “Judgment
System.”
The judgment result of each step is displayed on the screen during the interval and also deliv-
ered as a signal from the SIGNAL I/O connector. The buzzer does not sound.
Total judgment
When the program test is complete, a total judgment of all steps is displayed. If the judgment
of all steps is the same, the total judgment is also the same. If various judgments are mixed,
the judgment with the highest precedence becomes the total judgment. See the following
example.
ABORT setting and FAIL judgment
If ABORT is set to OFF in the example of program 01 above, the test will be executed to step
05, and the total judgment will be U-FAIL.
If ABORT is set to ON, CONTACT FAIL occurs in step 01, and the program test will stop at
this point. The total judgment will be CONTACT FAIL, and the test results up to step 01 can
be stored.
Storing the Test Result
The storage function is assigned to a function key while the judgment result is displayed. For
details, see “Storing the Test Results.”
Releasing the Judgment Result
Press the STOP switch to set the TOS3200 to READY status.
See
Step
Judgment of each step
Order of
prece-
dence
Judgment
Program 00
Program 01
1
U-FAIL
00
PASS
PASS
2
L-FAIL
01
PASS
CONTACT FAIL
3
CONTACT FAIL
02
PASS
U-FAIL
4
PASS
03
PASS
PASS
04
PASS
L-FAIL
05
PASS
PASS
Total
judgment
PASS
U-FAIL
See
Содержание TOS3200
Страница 15: ...General Description This chapter gives an overview and describes the features...
Страница 20: ...20 TOS3200 Options This page has been intentionally left blank...
Страница 29: ...Panel Control Basics This chapter explains how to select items on the screen and how to enter values...
Страница 38: ...38 TOS3200 Panel Memory This page has been intentionally left blank...
Страница 67: ...Meter Mode Measurement This chapter explains the touch current measurement in Meter Mode Meter Mode Measurement...
Страница 73: ...Program Test This chapter describes how to create and edit sequence pro grams and how to execute the test...
Страница 105: ...Specifications This chapter gives the specifications and external dimensions...
Страница 115: ...Appendix A A List of Default Settings B System Settings C Protection Function D Troubleshooting...
Страница 129: ......