TOS3200
103
Time Settings and Calibration Management
8
Maintenance
CAL. PROTECT (calibration due protection)
Sets the action taken when the calibration date expires.
If CAL. PROTECT is turned ON, the TOS3200 will display the blinking message “CAL DATE
EXPIRED” to notify that the calibration date is due. You cannot use the TOS3200 in this con-
dition.
Press the STOP switch to display SYSTEM5/5, and turn CAL. PROTECT off to continue
using the TOS3200.
If the calibration date is due and CAL. PROTECT is turned off, the message “CAL DATE
EXPIRED” will appear at power-on. You can continue using the TOS3200 by pressing the
STOP switch.
Calibration
The TOS3200 is shipped after carrying out appropriate calibrations. We recommend periodic
calibration to maintain the performance over an extended period.
To achieve periodic calibration, set an appropriate calibration due date (ALARM). For calibra-
tion, contact your Kikusui agent or distributor.
Содержание TOS3200
Страница 15: ...General Description This chapter gives an overview and describes the features...
Страница 20: ...20 TOS3200 Options This page has been intentionally left blank...
Страница 29: ...Panel Control Basics This chapter explains how to select items on the screen and how to enter values...
Страница 38: ...38 TOS3200 Panel Memory This page has been intentionally left blank...
Страница 67: ...Meter Mode Measurement This chapter explains the touch current measurement in Meter Mode Meter Mode Measurement...
Страница 73: ...Program Test This chapter describes how to create and edit sequence pro grams and how to execute the test...
Страница 105: ...Specifications This chapter gives the specifications and external dimensions...
Страница 115: ...Appendix A A List of Default Settings B System Settings C Protection Function D Troubleshooting...
Страница 129: ......