44
TOS3200
Connecting the EUT
Connecting the Test lead (TL21-TOS) (Cont’d)
The test lead connection varies depending on the type of TC to be measured and the EUT
class.
Table 4-1
Test lead connections for the type of TC measurement
Test lead A: Test lead connected to measurement terminal A
Test lead B: Test lead connected to measurement terminal B
If you select the PROBE item in the TC measurement screen (TC1/2), the LEDs located
above the measurement terminals will illuminate for those that require the test leads to be
connected.
Fig. 4-6 LED connection indications
Setting
Connection destination of the test lead
PROBE
COND
Class I equipment
Class 0I equipment
Class II equipment
E
NC
P
E
NORM
Connect test lead A to a part of the enclo-
sure that is not connected to the protective
ground [Connection a of Fig. 4-7].
Connect test lead A to the enclosure.
FLT
NEU
FLT
PE
Connect test lead A to a part of the enclo-
sure that is not connected to the protective
ground [Connection a of Fig. 4-7]
or a part that is connected to the protective
ground.
E
NC
E
NC
NORM
Connect test leads A and B to parts of the
enclosure that is not connected to the pro-
tective ground (two isolated locations)
[Connection b of Fig. 4-7].
Connect test leads A and B to the enclo-
sure (two isolated locations).
FLT
NEU
FLT
PE
Connect test leads A and B to parts of the
enclosure that is not connected to the pro-
tective ground (two isolated locations)
[Connection b of Fig. 4-7]
or to a part that is not connected to the
protective ground and a part that is con-
nected to the protective ground [Connec-
tion c of Fig. 4-7].
E
NC
L
IV
Connect test lead A to a part of the enclo-
sure that is not connected to the protective
ground [Connection a of Fig. 4-7].
Connect test lead A to the enclosure.
E
NC
N
EU
A
B
DANGER
LEAKAGE CURRENT
CAT II
REMOTE
Connect a test lead to
the measurement
terminal for which the
LED is illuminated.
Содержание TOS3200
Страница 15: ...General Description This chapter gives an overview and describes the features...
Страница 20: ...20 TOS3200 Options This page has been intentionally left blank...
Страница 29: ...Panel Control Basics This chapter explains how to select items on the screen and how to enter values...
Страница 38: ...38 TOS3200 Panel Memory This page has been intentionally left blank...
Страница 67: ...Meter Mode Measurement This chapter explains the touch current measurement in Meter Mode Meter Mode Measurement...
Страница 73: ...Program Test This chapter describes how to create and edit sequence pro grams and how to execute the test...
Страница 105: ...Specifications This chapter gives the specifications and external dimensions...
Страница 115: ...Appendix A A List of Default Settings B System Settings C Protection Function D Troubleshooting...
Страница 129: ......